EasyManuals Logo

Tektronix TDS1012 Programmer's Manual

Tektronix TDS1012
250 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #145 background imageLoading...
Page #145 background image
Command Descriptions
TDS200/1000/2000 Series Oscilloscope Programmer Manual
2−127
RS-232
RS232:HARDFlagging { ON | OFF | <NR1> }
RS232:HARDFlagging?
<ON> or <NR1> 0 activates hard flagging and turns off soft flagging.
<OFF> or <NR1> = 0 deactivates hard flagging (RTS always asserted).
RS232:HARDFlagging ON
activates hard flagging and deactivates soft flagging.
RS232:PARity
Sets or queries the parity used for all RS-232C data transfers. When
parity is odd or even, the oscilloscope generates the selected parity
on output and checks all input against the selected parity. When
parity is none, the oscilloscope performs no input parity error checks
and generates no output parity.
When the parity (ninth) bit does not match the parity type, the
oscilloscope reports a parity error. If another command is sent
immediately after this command, without waiting for the parity to be
programmed, the first few characters may be lost.
RS-232
RS232:PARity { EVEN | ODD | NONe }
RS232:PARity?
EVEN sets even parity.
Group
Syntax
Arguments
Examples
Group
Syntax
Arguments

Table of Contents

Other manuals for Tektronix TDS1012

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS1012 and is the answer not in the manual?

Tektronix TDS1012 Specifications

General IconGeneral
BrandTektronix
ModelTDS1012
CategoryTest Equipment
LanguageEnglish

Related product manuals