Command Descriptions
2−128
TDS200/1000/2000 Series Oscilloscope Programmer Manual
ODD sets odd parity.
NONe sets no parity (no ninth bit transmitted).
RS232:PARity EVEN
sets even parity.
RS232:SOFTFlagging
Sets or queries the input and output soft flagging over the RS-232C
port. After receiving an XOFF (DC3), the oscilloscope sends two or
less characters. The oscilloscope begins transmitting data again when
it receives an XON (DC1) character.
The oscilloscope sends an XOFF character when its input buffer is
running out of space. After sending an XOFF character it can receive
at least 20 more bytes. It sends XON when its input buffer has an
acceptable number of free bytes.
When soft flagging is enabled and binary data is transferred, data
transmission will lock up if the data contains XOFF or XON
characters.
Hard flagging and soft flagging cannot be active at the same time.
However, hard and soft flagging can both be inactive at the same
time. Activating soft flagging deactivates hard flagging. Activating
hard flagging deactivates soft flagging.
RS-232
RS232:SOFTFlagging { ON | OFF | <NR1> }
RS232:SOFTFlagging?
ON or NR1 ≠ 0 activates soft flagging and deactivates hard flagging.
OFF or NR1 = 0 deactivates soft flagging.
Examples
Group
Syntax
Arguments