TRIGger:MAIn:HOLDOff?
might return :TRIGGER:MAIN:HOLDOFF:VALUE 5.0E–7
TRIGger:MAIn:HOLDOff:VALue
Sets or queries the oscilloscope trigger holdoff value.
Trigger
TRIGger:MAIn:HOLDOff:VALue <NR3>
TRIGger:MAIn:HOLDOff:VALue?
<NR3> is the main trigger holdoff value, in the range of 500 ns
to 10 s.
TRIGger:MAIn:HOLDOff:VALue 10
sets the holdoff value to 10 s.
TRIGger:MAIn:LEVel
Sets the oscilloscope edge (all models) and pulse width (TDS1000
and TDS2000 series only) trigger level. This command is equivalent
to adjusting the front-panel TRIGGER LEVEL knob.
NOTE. When the edge trigger source is set to AC LINE, the
oscilloscope ignores the set form of the command and generates
event 221 (Settings conflict).
When the edge trigger source is set to AC LINE, the query form of
the command returns zero.