EasyManua.ls Logo

Tektronix TDS1012 - Page 162

Tektronix TDS1012
250 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Command Descriptions
2−144
TDS200/1000/2000 Series Oscilloscope Programmer Manual
Trigger
TRIGger:MAIn:LEVel <NR3>
TRIGger:MAIn:LEVel?
<NR3> the main trigger level, in volts.
TRIGger:MAIn:LEVel?
might return 1.4, indicating that the main edge trigger is set to
1.4 V.
TRIGger:MAIn:MODe
Sets or queries the trigger mode for the Edge (all models) and Pulse
width (TDS1000 and TDS2000 series only) trigger types.
Trigger
TRIGger:MAIn:MODe { AUTO | NORMal }
TRIGger:MAIn:MODe?
AUTO generates a trigger if a trigger is not detected within a specific
time period. AUTO also enables scan mode for sweep speeds of
100 ms/div and slower.
NORMal waits for a valid trigger event.
TRIGger:MAIn:MODe AUTO
specifies that a trigger event is automatically generated.
Group
Syntax
Arguments
Examples
Group
Syntax
Arguments
Examples

Table of Contents

Other manuals for Tektronix TDS1012

Related product manuals