Command Descriptions
2−144
TDS200/1000/2000 Series Oscilloscope Programmer Manual
Trigger
TRIGger:MAIn:LEVel <NR3>
TRIGger:MAIn:LEVel?
<NR3> the main trigger level, in volts.
TRIGger:MAIn:LEVel?
might return 1.4, indicating that the main edge trigger is set to
1.4 V.
TRIGger:MAIn:MODe
Sets or queries the trigger mode for the Edge (all models) and Pulse
width (TDS1000 and TDS2000 series only) trigger types.
Trigger
TRIGger:MAIn:MODe { AUTO | NORMal }
TRIGger:MAIn:MODe?
AUTO generates a trigger if a trigger is not detected within a specific
time period. AUTO also enables scan mode for sweep speeds of
100 ms/div and slower.
NORMal waits for a valid trigger event.
TRIGger:MAIn:MODe AUTO
specifies that a trigger event is automatically generated.