Command Descriptions
2−150
TDS200/1000/2000 Series Oscilloscope Programmer Manual
TRIGger:MAIn:VIDeo? (Query Only)
Returns the main video trigger parameters.
Trigger
TRIGger:MAIn:VIDeo?
TRIGger:MAIn:VIDeo?
might return :TRIGGER:MAIN:VIDEO:SOURCE CH1;SYNC FIELD;PO−
Larity NORMAL
TRIGger:MAIn:VIDeo:LINE
NOTE. You can use this command ONLY with the TDS1000 and
TDS2000 series.
Sets or queries the line number for the video trigger when
TRIGger:MAIn:VIDeo:SYNC is set to LINENUM. This is
equivalent to selecting a line number in the Trigger/Video menu.
Trigger
TRIGger:MAIn:VIDeo:LINE <NR1>
TRIGger:MAIn:VIDeo:LINE?
<NR1> specifies the line number. The minimum is 1. The maximum
is 525 for NTSC, 625 for PAL and SECAM.