Command Descriptions
TDS200/1000/2000 Series Oscilloscope Programmer Manual
2ā149
TRIGger:MAIn:PULse:WIDth:WIDth .000000123
followed by TRIGGER:MAIN:PULse:WIDth:WIDth? might return
1.155Eā7
TRIGger:MAIn:TYPe
Sets or queries the type of oscilloscope trigger. This is equivalent to
setting the Type option in the Trigger menu.
Trigger
TRIGger:MAIn:TYPe { EDGE | VIDeo | PULse }
TRIGger:MAIn:TYPe?
EDGE is a normal trigger. A trigger event occurs when a signal passes
through a specified voltage level in the specified direction and is
controlled by the TRIGger:MAIn:EDGE commands.
VIDeo specifies that a trigger occurs when a specified signal is found
and is controlled by the TRIGger:MAIn:VIDeo commands.
PULse specifies that a trigger occurs when the specified signal meets
the pulse width criteria that is controlled by the
TRIGger:MAIn:PULse commands.
TRIGger:MAIn:TYPe?
might return VIDEO indicating that the main trigger type is a video
trigger.
Examples
Group
Syntax
Arguments
Examples