Command Descriptions
2−160
TDS200/1000/2000 Series Oscilloscope Programmer Manual
RI specifies signed integer data-point representation.
RP specifies positive integer data-point representation.
WFMPre:BN_Fmt RP
specifies that the binary waveform data are positive integer
data-points.
WFMPre:BN_Fmt?
returns either RI or RP as the current waveform data format.
DATa:ENCdg
WFMPre:BYT_Nr
Sets or queries the data width for the waveform to be transferred.
This command is equivalent to DATa:WIDth. Changing
WFMPre:BYT_Nr also changes WFMPre:BIT_Nr and DATa:WIDth.
Waveform
WFMPre:BYT_Nr <NR1>
WFMPre:BYT_Nr?
<NR1> is an integer in the range of 1 to 2 that sets the number of
bytes per point.
WFMPre:BYT_Nr 2
specifies that there are 2 bytes per waveform data point.
Arguments
Examples
Related Commands
Group