Command Descriptions
2−136
TDS200/1000/2000 Series Oscilloscope Programmer Manual
<NR1> is an integer value in the range from 0 to 255. The binary bits
of the SRER are set according to this value. Using an out-of-range
value causes an execution error. The power-on default for SRER is 0
if *PSC is 1. If *PSC is 0, the SRER maintains its value through a
power cycle.
*SRE 48
sets the bits in the SRER to 00110000 binary.
*SRE?
might return a value of 32, showing that the bits in the SRER have
the binary value 00100000.
*CLS, DESE, *ESE, *ESR?, EVENT?, EVMSg?, FACtory, *PSC,
*STB?
*STB? (Query Only)
(Read Status Byte) query returns the contents of the Status Byte
Register (SBR) using the Master Summary Status (MSS) bit. Refer
to the Status and Events chapter on page 3-1 for more information.
Status and Error
*STB?
<NR1>
*STB?
might return the value 96, showing that the SBR contains the binary
value 01100000.
Arguments
Examples
Related Commands
Group
Syntax
Returns
Examples