Command Descriptions
TDS200/1000/2000 Series Oscilloscope Programmer Manual
2−71
<wfm> is the location of the waveform data that will be transferred
from the oscilloscope to the external device. Allowable values are
CH<x>, MATH, and REF<x>.
DATa:SOUrce REFB
specifies that reference waveform REFB will be transferred in the
next CURVe? query.
DATa:SOUrce?
might return REFA, indicating the source for the waveform data that
is transferred using CURVe?.
CURVe?, WFMPre?, WAVFrm?
DATa:STARt
Sets or queries the starting data point for waveform data transfers.
This command lets you transfer partial waveforms to and from the
oscilloscope.
Waveform
DATa:STARt <NR1>
DATa:STARt?
<NR1> is an integer value that ranges from 1 to 2500, and specifies
the first data point that will be transferred. Data is transferred from
<NR1> to DATa:STOP or 2500, whichever is less. When
DATa:STOP is less than DATa:STARt, the values are swapped
internally for CURVe?.
Arguments
Examples
Related Commands
Group
Syntax
Arguments

Table of Contents