Command Descriptions
2−40
TDS200/1000/2000 Series Oscilloscope Programmer Manual
ON | RUN | <NR1> ≠ 0 starts acquisition and display of waveforms. If
the command was issued in the middle of an acquisition sequence
(for instance averaging), RUN restarts the sequence, discarding any
data accumulated before the STOP. It also resets the number of
acquisitions.
ACQuire:STATE RUN
starts acquisition of waveform data and resets the number of
acquisitions count (NUMACq) to zero.
ACQuire:STATE?
returns 0 or 1, depending on whether or not the acquisition system is
running.
*OPC?
ACQuire:STOPAfter
Tells the oscilloscope when to stop taking acquisitions.
Acquisition
ACQuire:STOPAfter { RUNSTop | SEQuence}
ACQuire:STOPAfter?
RUNSTop specifies that the run and stop states should be determined
by pressing the front-panel RUN/STOP button or issuing the
ACQuire:STATE command.
Examples
Related Commands
Group
Syntax
Arguments