6. Check that the c
urrent output at J204 of the Display Adapter board for each
wire is around 25 mA. If the reading is not correct, the Display Adapter
module is probably defective. Replace it.
Figure 6-9: Measuring the backlight voltage
TDS2000C Series Oscilloscope Service Manual 6–33