Command Descriptions
TDS200/1000/2000 Series Oscilloscope Programmer Manual
2−97
HORizontal:DELay:SCAle 9.0E–6
sets the window scale to 10 ms per division. Since 9 ms is not a
correct value within the 1–2.5–5 sequence, it is automatically set to
the closest correct value.
HORizontal:DELay:SCAle?
might return 1.0E–3, indicating that the window scale is 1 ms per
division.
HORizontal:DELay:SECdiv
This command is identical to the HORizontal:DELay:SCAle. It is
provided to maintain program compatibility with some older models
of Tektronix oscilloscopes.
HORizontal:MAIn? (Query Only)
Returns all settings for the oscilloscope main time base. The
HORizontal:MAIn:SECdiv and HORizontal:MAIn:SCAle com-
mands are identical so only HORizontal:MAIn:SCAle is returned.
Horizontal
HORizontal:MAIn?
All settings for the main time base.
HORizontal:MAIn?
might return :HORIZONTAL:MAIN:POSITION 0.0E0;SCALE 5.0E–4
Group
Syntax
Returns
Examples