Command Descriptions
TDS200/1000/2000 Series Oscilloscope Programmer Manual
2−67
DATa
Sets or queries the format and location of the waveform data that is
transferred with the CURVe command. Since DATa:DESTination and
DATa:TARget are equivalent, only DATa:DESTination is returned by
the DATa? query.
Waveform
DATa { INIT }
DATa?
INIT initializes the waveform data parameters to their factory
defaults.
DATa INIT
initializes the waveform data parameters to their factory defaults:
DESTINATION = REFA (=TARGET)
ENCDG = RIBINARY
SOUrce = CH1
START = 1
STOP = 2500
WIDTH = 1
DATa?
might return the string :DATA:ENCDG RPBINARY;
DESTINATION REFA; SOURCE REFB;START 1;STOP 500;WIDTH 2
CURVe, WAVFrm
Group
Syntax
Arguments
Examples
Related Commands