TDS1000 and TDS2000 Series Digital Storage Oscilloscopes Service Manual
6-- 29
Installation. To install the switch keypad, refer to Figure 6-- 18 and follow these
steps:
1. Place the front case face down on a soft surface.
2. Align the switch keypad into the front case using the guide posts and the
holes in the keypad.
CAUTION. Keep the electrical contacts on the back of the switch pad clean. Skin
oils and dust will inhibit good electrical contact.
3. Check that all buttons are properly seated into place before proceeding.
4. Place the entire internal assembly into the front case.
5. Use the installation procedure for each module removed to reassemble the
oscilloscope.