Appendix A: Specificati ons
TDS1000/2000-Series Digital Oscilloscope User Manual
157
Oscilloscope Specifications (Cont.)
Trigger
Trigger Sensitivity,
E
T
i
T
Coupling Sensitivity
y
,
Edge Trigger Type
DC
CH1, CH2,
CH3, CH4
1 div from DC to 10 MHz*,
1.5 div from 10 MHz* to Full
EXT 200 mV from DC to 100 MHz*,
350 mV from 100 MHz to
200 MHz*
EXT/5 1 V from DC to 100 MHz*,
1.5 V from 100 MHz to 200 MHz*
Trigger Sensitivity,
E
T
i
T
Coupling Sensitivity
y
,
Edge Trigger Type,
t
y
i
c
l
AC Same as DC at 50 Hz and above
t
y
i
c
l
NOISE
REJ
Reduces the DC-coupled trigger sensitivity by
2timesfor>10mv/divto5V/div
HF REJ Same as the DC-coupled limit from DC to 7 kHz,
attenuates signals above 80 kHz
LF REJ Same as the DC-coupled limits for frequencies
above 300 kHz, attenuates signals below 300 kHz
Trigger Level Range Source Range
CH1, CH2,
CH3, CH4
±8 divisions from center of screen
EXT ±1.6 V
EXT/5 ±8V
* Bandwidth reduced to 6 MHz with a 1X probe.

Table of Contents

Other manuals for Tektronix TDS2012

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS2012 and is the answer not in the manual?

Tektronix TDS2012 Specifications

General IconGeneral
BrandTektronix
ModelTDS2012
CategoryTest Equipment
LanguageEnglish

Related product manuals