Command Descriptions
TDS200/1000/2000 Series Oscilloscope Programmer Manual
2−173
WFMPre, Additional Commands for Compatibility
The set form of these additional commands is ignored. The query
form generates event messages 100 (Command Error) and 420
(Query Unterminated). These commands are included for compati-
bility purposes only.
Table 2-26 lists additional WFMPre commands.
Table 2−26: Additional WFMPre commands
Command Argument Description
WFMPre:XMUlt <NR3> Horizontal (X−axis) scale factor
WFMPre:XOFf <NR3> Horizontal (X−axis) offset
WFMPre:ZMUlt <NR3> Z−axis scale factor
WFMPre:ZOFf <NR3> Z−axis offset
WFMPre:ZUNit <QString> Z−axis units
WFMPre:ZZEro <NR3> Z−axis origin offset
WFMPre:<wfm>? (Query Only)
Returns the waveform formatting data for the waveform specified by
<wfm>, if that waveform is active or displayed. If that waveform is
not active or displayed, the query fails and the oscilloscope generates
event message 420 (Query unterminated) and 2244 (waveform
requested is not active).
Waveform
Group

Table of Contents

Other manuals for Tektronix TDS2022

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS2022 and is the answer not in the manual?

Tektronix TDS2022 Specifications

General IconGeneral
BrandTektronix
ModelTDS2022
CategoryTest Equipment
LanguageEnglish

Related product manuals