Command Descriptions
2−176
TDS200/1000/2000 Series Oscilloscope Programmer Manual
WFMPre:<wfm>:XUNit
Same as WFMPre:XUNit, except that <wfm> specifies the
source/destinaton waveform instead of DATa:SOUrce and
DATa:DESTINATION.
WFMPre:<wfm>:XZEro
Same as WFMPre:XZEro, except that <wfm> specifies the
source/destinaton waveform instead of DATa:SOUrce and
DATa:DESTINATION.
For set commands, if <wfm> is not a reference waveform, the
oscilloscope generates error 2241.
WFMPre:<wfm>:YMUlt
Same as WFMPre:YMUlt, except that <wfm> specifies the
source/destinaton waveform instead of DATa:SOUrce and
DATa:DESTINATION.
For set commands, if <wfm> is not a reference waveform, the
oscilloscope generates error 2241.
WFMPre:<wfm>:YOFf
Same as WFMPre:YOFf, except that <wfm> specifies the source/
destinaton waveform instead of DATa:SOUrce and DATa:DESTINA-
TION.
For set commands, if <wfm> is not a reference waveform, the
oscilloscope generates error 2241.