empower your supply chain.
V1.0-11/2022-EN · Technical Changes and Errors excepted 15
MainTechnicalSpecicationstempmate.®-S2
T
RecordingOptions Single-Use
Dimension [mm] 93 x 47 x 8 mm
Weight[g] 30g
Casing
EcofriendlyHardcase
Battery CR2450 lithium metal button cell battery
Connection Interface
USB2.0,A-Type(integrated)
Display Multifunction LCD
Protection Class IP65
Shelf Life 24 months
TemperatureRange –30 °C to 70 °C
Temperature Accuracy
±0.3°C(-20to40°C)±0.5(others)
Temperature Resolution 0.1 °C
HumidityRange n/a
HumidityAccuracy n/a
HumidityResolution n/a
AmbientLightSensor Yes
MemoryCapacity(measurements) 32,000points(T)
RunTime(LoggingInterval) Upto180days(10min.)=StandardModel
Other models only on request
Data Export
PDF&CSV
AlarmConguration upto5pointscustomizable,singleorcumulative
Startup Mode Button
Stop Mode Button or by Software
OtherSpecications
TemporaryPDF ThePDFReportisavailableatanytime
withoutstoppingtheLogger
Software PDFreader+optionaltempbase2Software
Reprogrammable Yes,withoptionaltempbase2Software
Backup-Log-Feature Yes
Custom ID and Travel Description Yes
CalibrationCerticate Yes,with6points(hardcopy)&PDFinmemory
Conformity EN12830,CE,RoHS
SerialNumberIdentication Yes
MainTechnicalSpecicationstempmate.®-S2
TH
RecordingOptions Single-Use
Dimension [mm] 93 x 47 x 8 mm
Weight[g] 30g
Casing
EcofriendlyHardcase
Battery CR2450 lithium metal button cell battery
Connection Interface
USB2.0,A-Type(integrated)
Display Multifunction LCD
Protection Class IP52
Shelf Life 24 months
TemperatureRange –30 °C to 70 °C
Temperature Accuracy
±0.3°C(-20to40°C)±0.5(others)
Temperature Resolution 0.1 °C
HumidityRange 0-100%rH
HumidityAccuracy ±3%rH(20to80%rH)±5%(others,at25°C)
HumidityResolution 0.1%rH
AmbientLightSensor Yes
MemoryCapacity(measurements) 32,000points(T/rH)
RunTime(LoggingInterval) Upto180days(10min.)=StandardModel
Other models only on request
Data Export
PDF&CSV
AlarmConguration up to 5 points temperature and 2 points humidity customizable
Startup Mode Button
Stop Mode Button or by Software
OtherSpecications
TemporaryPDF ThePDFReportisavailableatanytime
withoutstoppingtheLogger
Software PDFreader+optionaltempbase2Software
Reprogrammable Yes,withoptionaltempbase2Software
Backup-Log-Feature Yes
Custom ID and Travel Description Yes
CalibrationCerticate Yes,with6points(hardcopy)&PDFinmemory
Conformity EN12830,CE,RoHS
SerialNumberIdentication Yes
Sinceenvironmentalinuencescanaffectthelifespanofbatteries,werecommendthatallourproductsbestoredintheconditions
thatareoptimalforthem.Inthisway,theshelflifegivenatthetimeofpurchasecanbeguaranteed.
Optimal storage conditions: +15°C to +23°C / 45% to 75% rH.