UNLOCKING POTENTIAL | TERUMOBCT.COM Page 10 of 15 Version 1.0 Rev 02/2021
©2021 Terumo BCT, Inc.
3. Installation
This section involves unpacking, temperature requirements, and installation of the device.
Unpacking and Inspection
1. Visually inspect the cardboard box for damage. Report any damage immediately.
2. Lift the device out of the cardboard box and place it on a flat surface.
The device is shipped in one cardboard box that includes:
1 Sealing unit
2 Power cords
1 Operator’s manual
1 Operator’s manual in other languages: on CD-ROM/USB stick
3. The above list is subject to change; refer to the packing list for an accurate description of contents.
If any parts are missing or if the parts are damaged, report it immediately.
4. Please keep all shipping and packaging materials, as they may be required for later transportation, at least during the
warranty period.
Environmental Requirements
To keep the device operating at its best, please observe the following:
Place the device on a flat surface free from dust, solvents, and acidic vapor.
Use the instrument in a dry area free from vibration and with a room temperature of 0 °C to 35 °C
(32 °F to 95 °F) and relative humidity 10% to 90% non-condensing.
Handle the components with care in a clean environment.
3.1 Installation Procedure
Stand-Alone Setup
Preparing T-SEAL III for use.
1. Unfold the front and back support wings.
2. Place T-SEAL III on a flat surface, out of direct sunlight.
3. Connect the power cord to the receptacle at the rear of the device.
4. Plug the power cord into a grounded outlet.
5. Perform a test seal on an empty or water-filled tube to ensure proper operation.
Cascade Setup
Warning: This is a Class A product when in cascade mode. In a domestic environment, this product may cause radio
interference, in which case the user may be required to take actions.
1. Unfold the positioning wing.
2. Fix the tube segment length by positioning the T-SEAL III devices in the correct slot of the positioning wing. The
segment length can be modified in increments of 1 cm. Minimum segment length is 7 cm (first slot position);
maximum segment length is 11 cm (last slot position)
3. The leader can be placed on the left side or the right side. The leader is selected by inserting a communication
termination plug into its COM IN. The last follower should have a communication termination plug into its COM
OUT.
Note: In a cascade setup with mixed T-SEAL II and T-SEAL III devices, the leader must be a T-SEAL III device.