Frequency for acoustic testing:
8000Hz
.
Warm-up time:
2min
Sampling interval:
Bar graph indication 125 ms approx.
Numeric indication 1 sec approx.
Data record capacity:
Data can be stored in the memory.
Max. 32000 data can be stored.
Max. 255 blocks can be split.
Display LCD
Display screens
:
4 digit numerical indication of sound level, from 30.0 to 130.0dB with 0.1dB
resolution.
Bar-graph indication of current sound level with 1dB resolution.
Sound level range indicator: 30–90dB, 40–100dB, 50–110dB, 60–120dB or
70–130dB in five ranges.
Time display; year – month – day and hour: minute: second.
Display update rate:
1 second
Display first indication:
Depends on the condition the instrument when it was last
turned off.
Warning indications:
Out-of-range indications:
OVER displayed at upper limit of the range
UNDER displayed at lower limit of the range
Outputs
AC output
(using selected frequency weighting)
Output voltage:
1Vrms (at full-scale of the range)
Output impedance:
5k
Ω
Load impedance:
1M
Ω
DC output
Output voltage:
10mV/dB
Output impedance:
5k
Ω
Load impedance:
1M
Ω
I/O connector:
Sound level meter control from and data output to a computer (RS232)
Clock:
Real–time (with calendar)
Power requirements
Qty 4 x 1.5V IEC R6P (size “AA”) manganese super heavy duty batteries or
equivalent.
Battery life
: Approx. 50 hours