Chapter 6 – Specifications
TestEquity 106 & 107 Temperature Chamber Page 6-1
Chapter 6 – Specifications
Model 106 & 107 Chamber Specifications
Temperature Range –42°C to +130°C
Control Tolerance ±0.5°C, ±0.2°C Typical (Measured at the control sensor after stabilization)
Uniformity ±1.0°C, ±0.5°C Typical (Variations throughout the chamber after stabilization)
Live Load Capacity @ +23°C 0°C –30°C –40°C
(Watts) 200 W 155 W 100 W 35 W
Heat Up Transition Time 8.5°C/minute (empty chamber, typical)
Cool Down Transition Time (empty chamber, typical)
End Temp
Start Temp to
+23°C 0°C –10°C –20°C –30°C –35°C –40°C
+23°C ----- 7 min 11 min 15 min 19 min 23 min 30 min
+85°C 13 min 20 min 24 min 28 min 32 min 36 min 45 min
*Note: Transition times are measured after a 30 minute soak at the start temperature. To calculate rate of change for
a particular condition, take the difference between the Start Temp and End Temp and divide by the Transition Time.
Cool Down Example (empty): From +85°C to –20°C = 105°C / 28 min = 3.75°C/min.
Power Requirements
Input Voltage 120 VAC nominal (110 to 126 VAC), Single Phase, 60 Hz
Current Draw 10 A maximum; Recommended Service 15 A
Heat of Rejection 3,000 BTUH
(maximum rated chamber load at maximum cooling rate from high temperature soak)
Workspace Dimensions Model 106: 9" W x 9" H x 13.25" D (0.62 cubic feet)
Model 107: 12" W x 9" H x 11.25" D (0.7 cubic feet)
Outside Dimensions Model 106: 23" W x 26" H x 16" D (nominal)
Model 107: 16.5" W x 26" H x 23" D (nominal), Door latch adds 2" to width.
Min. Installed Clearance 12" from the rear
Access Ports Model 106: 4" (2.83" ID) Port on left side, Supplied with foam plug
Model 107: 3" (3.83" ID) Port on left and right side (two total), Supplied with foam
plugs
Weight 124 pounds
Sound Level 52 dBA in cooling mode
(A-weighted, measured 36" from the front surface, 63" from the floor, in a free-standing environment)
NOTE: Performance is typical and based on operation at 23°C (73°F) ambient and nominal input voltage. This
product is designed for use in a normal conditioned laboratory. Operation at higher ambient temperatures will result
in decreased cooling performance. Low end limit derates to -38°C when operating above 27°C (80°F) ambient.
Operation above 30°C (85°F) or below 16°C (60°F) ambient is not recommended.