Product description14
Product description
At a glance
Protection cap: Park position
Differential pressure sensor
nipple connection
Display
Control keys
Battery compartment, holding
magnets (on rear)
First steps
²
Inserting bbatteries:
1 To open the battery compartment, push the battery cover
down.
2 Insert batteries (2x 1.5 V type AAA). Observe the polarity!
3 To close the battery compartment, push the battery cover
back on.
²
Basic ssettings ((configuration mmode):
Adjustable functions
· Unit of pressure/velocity: hPa, mbar, Pa, mmH2O, mmHg, psi, inH2O, inHG,
m/s, fpm
· Only if a unit of velocity is selected:
Density of the medium (Density): Adjust flashing numbers with , change to
the next number with .
· Auto off function: OFF, ON (instrument switches off automatically if no key is
pressed for 10 minutes)
1 When switching the instrument on, press and hold until
and appear on the display (configuration mode).
- The adjustable function is displayed. The current setting
flashes.
2 Press ( ) several times until the desired setting flashes.
3 Press ( ) to confirm the input.
4 Repeat steps 2 and 3 for all functions.
- The instrument changes to measuring mode.