27 | P a g e
Part Number - 00824 Revision K, DCO 2017-0074
© 2017 Thermedx
®
, LLC
Perforation Alarm (optionally enabled)
The perforation alarm can be enabled when
using deficit monitoring; it is off by default.
If enabled the alarm will sound when the rate
of deficit rise is faster than the value selected
(Default of 200ml/min up to flow rate limit for
the selected procedure).
Note: False positives can occur during periods
of high flow, such as when filling cavities, or
when removing the scope from a cavity
without closing inflow valve.
Print Summary
User may Print a summary of pertinent case
information such as current volume and
deficit.
Volume/Deficit snapshot
14:58
Fluid pumped (input) 30 ml
Fluid returned (output) 30 ml
Fluid deficit: 0 ml
Procedure ongoing…
Alarm and bag chime sound level
Alarm and bag chime sound level can be
adjusted by selecting the Adjustment Icon