ALARM, ERROR AND WARNING MESSAGES
WHERE “CALL FOR SUPPORT” IS INDICATED, PLEASE CALL YOUR DISTRIBUTOR OR THERMEDX
®
AT 888-542-
9276. FOR CUSTOMERS OUTSIDE THE US, PLEASE CALL YOUR DISTRIBUTOR OR THERMEDX
®
AT
011.1.440.542.0883.
Temperature, pressure, flow rate, and deficit
values exceeding limits may be accompanied
by an audible alarm.
Users may Mute audio from low level “set-
point” alarms. These alarms may also be
bypassed by touching the notification on
screen. Bypassing allows the user to continue
the procedure unless an upper “safety” limit
is reached.
Temperature has exceeded set-point by 3
degrees C. This can be caused by abrupt
changes in flow rates. User may press to
continue operation. System will not warm
until temperature has dropped below limit.
Temperature has exceeded 46 degrees C.
Wait for fluid to cool, system will then
continue operation.
Temperature has exceeded 46 degrees C,
and secondary temperature protection means
have engaged. Replace tube set and reboot
system to continue.
Pressure has exceeded set-point by 10% or
12mm/Hg (whichever is greater). This can be
caused by abrupt changes in flow rates. User
may bypass alarm, system will continue to
attempt to control pressure to the set-point.
Pressure has exceeded the maximum allowed
(varies by procedure). Fluid pressure must
drop to continue. The pump will reverse and
attempt to lower pressure.
Flow rate has exceeded set-point by 10%
Flow rate has exceeded maximum allowed
(varies by procedure).
Deficit has exceeded user selected set-point.
User may press to bypass, collect fluids to
lower deficit, or raise set-point.
Deficit has exceeded maximum limit. This
varies by procedure. Collect fluids to lower
deficit or end case.