What to do if I get a Carousel Error on my Thermo Scientific PrintMate A79510100 Laboratory Equipment?
T
Timothy NicholsAug 1, 2025
If you encounter a Carousel Error with your Thermo Scientific Laboratory Equipment, start by checking for any obstructions that might be preventing the carousel from rotating. Remove any obstructions if found. If this doesn't solve the problem, ensure that the hoppers are correctly seated on the carousel. Another potential cause is the dispense lever interfering with the hopper slider; in this case, remove the hopper at the output position, perform a dispense cycle, and then replace the hopper. You may also remove all hoppers to check for hidden obstructions. If the problem persists, it may be an instrument malfunction.