Installation and Operation Manual
Thermo Scientific Revco
®
General Purpose, Flammable Storage,
and Explosion Proof
Laboratory Refrigerators and Freezers

Questions and Answers

  • M
    Mark DouglasJul 29, 2025
    What to do if my Thermo Scientific REVCO Refrigerator unit warms up?
    • R
      Renee HarrisonJul 29, 2025
      If your Thermo Scientific Refrigerator is warming up, several factors could be the cause. First, ensure that the door is completely closed. If the door is closed, inspect the door seal, referring to Section 4.4 for guidance. If you recently loaded warm items, give the unit sufficient time to cool down. Check the power supply for proper voltage, and if there's no voltage, consult an electrician. Finally, if the compressor isn't running, have an electrician verify the voltage. If the compressor is running, contact an authorized service provider or the technical support hotline.
  • S
    Sherry CannonAug 8, 2025
    How to troubleshoot a Thermo Scientific REVCO that does not operate?
    • D
      Danielle OliverAug 9, 2025
      If your Thermo Scientific Refrigerator is not operating, start by checking that the cord is securely plugged in. Next, test the outlet by plugging in another appliance to see if it's working. If the outlet is dead, check the circuit breaker or fuses.
  • D
    Danielle HammondAug 18, 2025
    Why does the temperature fluctuate in my Thermo Scientific Refrigerator?
    • L
      Larry HarrisonAug 18, 2025
      Temperature fluctuations in your Thermo Scientific Refrigerator can be due to a couple of reasons. Begin by verifying that the cold control is correctly set. Also, ensure that the condenser is clean.

Summary

Introduction

Safety Considerations

Unpacking and Inspection

Installation

Location

Install the unit in a level area free from vibration and ambient heat.

Wiring

Connect the equipment to the correct power source and ensure proper grounding.

Leveling

Rotate leveling screws to ensure the unit is level and shim if necessary.

Door Seal

Check the door seal integrity by pulling a paper strip from between the door and gasket.

Operation

Temperature Control

Adjust cold controls to set the desired temperature, waiting 24 hours between adjustments.

Combination Refrigerator;Freezers

Understand temperature sensor location for combined units.

Defrosting

Turn off unit to melt frost exceeding 1/4 inch; do not use sharp objects.

Cleaning

Troubleshooting

Flammable Storage Units

Safety and Usage Standards

Follow guidelines for Class I, Division 2 storage of flammable liquids.

Electrical Requirements

Ensure proper grounding and use appropriate wiring for flammable storage units.

Explosion-Proof Units

Safety and Usage Standards

Approved for Class I, Division 1 locations; review flammable storage guidelines.

Electrical Requirements

Install using rigid metal conduit and a disconnect switch suitable for hazardous locations.

Thermo Scientific REVCO Series Specifications

General IconGeneral
DefrostManual
ControlMicroprocessor
Alarm SystemAudible and visual alarms for temperature deviations
MaterialStainless steel or coated steel exterior
ApplicationLaboratory, pharmaceutical
Voltage115V or 230V (depending on model)
ShelvesAdjustable shelves
Door LockYes
CertificationsUL
RefrigerantCFC-free

Related product manuals