3.6 Setup (Configuration)
Select mode, store and exit: Press simultaneously for >1 sec
Parameter Display Settings Description
example bold = default
SET ALTI 4321 m +/- (adjust) 1m/1ft steps, increasing to 10m/10ft steps Adjustment see chap. 3.7
SET QNH 1013.2 hPa +/- (adjust) 0.1mbar steps, increasing to 1mbar steps
SET DATE/TIME 01.02.97 +/- (adjust) Set day/month/year, minutes/hours Adjustable value
12:34:00
flashes
Next value
Save setting
SET Unit Pressure hPa mbar hPa mmHg inH
2
O inHg psia Units of measurement
SET Unit Temp. °C °C °F
SET Unit Humidity %rF %rF %rH
SET Unit Altitude m m ft
SET Unit Tendency ../minute ../h ../min
SET Rec. Interval 1 s (16 min) manual 1, 5, 10, 20, 30 sec Interval time datalogging
1, 2, 5, 10, 20, 30 min (max. recording time)
1, 3, 6, 24 h
SET Baudrate 9600 9600 2400 1200 Data transfer rate
SET Timeout 1 min continuous 1 min 10 min 30 min 60 min autom. cut-out time
SET Mixed Mode BARO BARO QNH ALTITUDE Adjust mixed mode
Reset RESET ? [CLEAR] General reset to
press CLEAR works setting
Print Setup PRINT ? [START] Log settings
press START
At Print Setup
Press > 5 sec
SET BARO-OFFSET +1.2 hPa +/- (adjust) only single 0.1 mbar steps possible Adjustment see chap. 3.7
Influences measuring accuracy
At SET BARO-OFFSET
Press > 5 sec
SET TEMP2-OFFSET -0.3 °C +/- (adjust) only single 0.1 °C steps possible Adjustment see chap. 3.7
Influences measuring accuracy
''
''
'
''
''
'
''
''
'
''
''
'