EasyManua.ls Logo

Thommen HM30 - Page 22

Default Icon
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
22
remote * 182 CR (tab)"ok" * 13 CR Switch to remote control / switch on
local * 53 CR (tab)"ok" * 13 CR Switch over to keypad /
exit remote control
off * 101 CR (tab)"ok" * 13 CR Switch off instrument
readall * 255 CR (tab)BARO_"value"_"unit"_QNH_"value"_"unit".. * Checksum CR Interrogate all meas. values simultaneously
readbaro * 106 CR (tab)"value"_"unit"_ * Checksum CR Interrogate current pressure measured value
readqnh * 13 CR Interrogate QNH measured value
readhumid * 221 CR Interrogate humidity measured value
readtemp1 * 173 CR Interrogate temperature 1 measured value
readdew * 6 CR Interrogate dew-point
readtemp2 * 174 CR Interrogate temperature 2 measured value
readtempint * 199 CR Interrogate internal temperature (±4°C)
readalti * 112 CR Interrogate altitude measured value
readfast * 116 CR (tab)"value"_*checksum CR (tab)"value"* ... Fast measured value interrogation of the
previous parameter (previous command
e.g. "readtemp1")
$ (tab)"ok" * 13 CR Exit fast measured value interrogation
clearmem * 112 CR (tab)"ok" * 13 CR Clear datalogging memory
readrecord * 69 CR (tab)"Date"_"Time"_"Recinterval"_ * Checksum CR Read out stored datalogging data
(tab)"type of measurement"["unit"]_ * Checksum CR
(tab)"value"_ * Checksum CR
(tab)"record_stopped"_ * 241 CR Recording stopped
(tab)"out_of_range"_ * 205 CR Measurement out of range
(tab)"record_end"_ * 41 CR Recording ended
readsetup * 247 CR (tab)Code (see decoding in table page 12) * Checksum CR Readout instantaneous configuration
readbat * 253 CR (tab)"voltage value"_V_("full"/"empty") * Checksum CR Interrogate battery voltage (V)
setqnh_"value" * Chk CR (tab)"ok" * 13 CR Set QNH
setalti_"value" * Chk CR (tab)"ok" * 13 CR Set present location altitude
setunit_hpa * 143 CR (tab)"ok" * 13 CR Select pressure units
setunit_mbar * 248 CR
setunit_mmhg * 255 CR
setunit_inh2o * 54 CR
Control commands/Syntax Reply from HM30/Syntax Description

Table of Contents