EasyManuals Logo

THORLABS GVS011 User Manual

THORLABS GVS011
40 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #27 background imageLoading...
Page #27 background image
Rev 19 Oct 2020
Appendix A Specifications and Associated Parts
Page 25
Notes
b Measured after leaving the unit powered up for at least 8 hours (max drift is <0.8
mrad in 8hrs upon power on)
Parameter Value
Motor & Position Sensor
Linearity(50% Full Travel) 99.9%
Scale Drift
b
<200 PPM/°C(Max)
Zero Drift
b
<20 µRad/°C(Max)
Repeatability 15 µRad
Resolution
With GPS011 Linear PSU
With standard switch mode PSU
0.00086 ° (15 µRad)
0.004° (70 µRad)
Average Current 1 A
Peak Current 10 A
Load Mirror Aperture 10 mm
Maximum Scan Angle (Mechanical Angle) ±20°
(with 0.5V/° scaling factor)
Motor Weight (inc cables, excl bracket) 94 g
Recommended Operating Temperature Range 15 to 35° C
Position Sensor Output Range 40 to 80 µA

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the THORLABS GVS011 and is the answer not in the manual?

THORLABS GVS011 Specifications

General IconGeneral
BrandTHORLABS
ModelGVS011
CategoryScanner
LanguageEnglish

Related product manuals