EasyManua.ls Logo

Thyracont VD85 - Error Messages and Malfunction

Thyracont VD85
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
40
vd84-201113
Error messages and malfunction
Problem
Possible Cause
Correction
high measurement error
contamination, ageing, ex-
treme temperature, malad-
justment
readjustment
display shows "ur"
pressure under range
(pressure < 1x10
-3
mbar)
error message "Err"
adjustment done at wrong
pressure
measurement error out of
adjustment range
displayed pressure must be
>40mbar f. atmosphere ad-
justment, <4x10
-2
mbar f.
zero adjustment
send unit for repair
error message "Err1"
sensor not connected
defective sensor
check connection and plug
send unit for repair