MANUFACTURER’S DECLARATION - ELECTROMAGNETIC EMISSIONS
The digital ear thermometer is intended for use in the electromagnetic environment (for home
healthcare)speciedbelow.Thecustomerortheuserofthedigitalearthermometershouldassure
that it is used in such an environment.
RF emissions
CISPR 11
RF emissions
CISPR 11
Group 1
Class B
The digital ear thermometer uses RF energy only for its
internal function. Therefore, its RF emissions are very
low and are not likely to cause any interference in nearby
electronic equipment.
The digital ear thermometer is suitable for use in all
establishments, including domestic establishments and
those directly connected to the public low-voltage power
supply network that supplies buildings used for domestic
purposes.
Emissions test Compliance Electromagnetic environment-guidance
(for home healthcare environment)
MANUFACTURER’S DECLARATION-ELECTROMAGNETIC IMMUNITY
TheDigitalEarThermometerisintendedforuseintheelectromagneticenvironmentspeciedbelow.The
customerortheuseroftheDigitalEarThermometershouldassurethatitisusedinsuchanenvironment.
NOTE 1At80MHzand800MHz,theseparationdistanceforthehigherfrequencyrangeapplies.
NOTE 2Theseguidelinesmaynotapplyinallsituations.Electromagneticpropagationisaffectedbyabsorptionandreectionfrom
structures, objects and people.
a.Fieldstrengthsfromxedtransmitters,suchasbasestationsforradio(cellular/cordless)telephonesandlandmobileradios,amateur
radio,AMandFMradiobroadcastandTVbroadcastcannotbepredictedtheoreticallywithaccuracy.Toassesstheelectromagnetic
environmentduetoxedRFtransmitters,anelectromagneticsitesurveyshouldbeconsidered.Ifthemeasuredeldstrengthinthe
locationinwhichtheDigitalEarThermometerisusedexceedstheapplicableRFcompliancelevelabove,theDigitalEarThermometer
should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as
reorientingorrelocatingtheDigitalEarThermometer.
Radiated RF
IEC 61000-4-3
10V/m
80MHz–2,7GHz
80% AM
at1kHz
10V/m
80MHz–2,7
GHz80%AM
at1kHz
Recommended separation distance
d=1,2√P
d=1, 2P80MHzto800MHz
d=2,3√P800MHzto2,7GHz
where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
distance in meters (m).
FieldstrengthsfromxedRFtransmitters,as
determined by an electromagnetic site survey, a should
be less than the compliance level in each frequency
range b. Interference may occur in the
vicinity of equipment marked with
the following symbol:
Immunity test
IEC 60601
test level
Compliance
level
Electromagnetic environment –
guidance