EasyManua.ls Logo

Tommee Tippee Digital ear thermometer - RF Separation Distance Guidance

Tommee Tippee Digital ear thermometer
20 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
MANUFACTURER’S DECLARATION – ELECTROMAGNETIC IMMUNITY
The digital ear thermometer is intended for use in the electromagnetic environment (for home healthcare)
speciedbelow.Thecustomerortheuserofthedigitalearthermometershouldassurethatitisusedin
such an environment.
Electrostatic
discharge(ESD)
IEC 61000-4-2
Power frequency
(50/60Hz)
magneticeld
IEC 61000-4-8
Contact±8kV
Ai2kV,4kV,±8
kV,±15kV
30 A/m
50Hzor
60Hz
Contact±8kV
Ai2kV,4kV,±8
kV,±15kV
30 A/m
50Hzor
60Hz
Floors should be wood, concrete or ceramic
tile.Ifoorsarecoveredwithsynthetic
material, the relative humidity should be at
least 30 %.
The digital ear thermometer power
frequencymagneticeldsshouldbeatlevels
characteristic of a typical location in a typical
home healthcare environment.
Immunity test
IEC 60601
test level
Compliance
level
Electromagnetic environment- guidance
(for home healthcare environment)
3 A/m
RECOMMENDED SEPARATION DISTANCE BETWEEN PORTABLE AND MOBILE RF
COMMUNICATIONS EQUIPMENT AND THE DIGITAL EAR THERMOMETER
The digital ear thermometer is intended for use in an electromagnetic environment (for home healthcare)
in which radiated RF disturbances are controlled. The customer or the user of the digital ear thermometer
can help prevent electromagnetic interference by maintaining a minimum distance between portable and
mobile RF communications equipment (transmitters) and the digital ear thermometer as recommended
below, according to the maximum output power of the communications equipment.
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be
estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1At80MHzand800MHz,theseparationdistanceforthehigherfrequencyrangeapplies.
NOTE 2 Theseguidelinesmaynotapplyinallsituations.Electromagneticpropagationisaffectedbyabsorptionandreectionfrom
structures, objects and people.
800MHzto2,7GHz
d=2,3√P
0,23
0,73
2,3
7,3
23
80MHzto800MHz
d=1,2√P
0,12
0,38
1,2
3,8
12
0,01
0,1
1
10
100
Rated maximum output
power of transmitter W
Separation distance according to frequency of transmitter m