EasyManua.ls Logo

TPS BLUE M ESP400 - Exhaust Connection

TPS BLUE M ESP400
36 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
BLUE M
Model ESP400 Industrial Oven
Page 14
5.4 Exhaust Connection
A 3” diameter exhaust port is located on the left side of the oven with a manual butterfly damper. An exhaust
guard is installed to prevent contact with the extremely hot port surface.
A draft diverter, vacuum breaker, exhaust hood, or similar device should be connected between the oven
and the exhaust. This connection prevents “chimney effect” which sucks heat out of the chamber and results
in slow run-up times or poor temperature uniformity. Never use an exhaust stack directly between the oven
and the exhaust.
Vent ductwork should be run to a location outside of the building (as necessary). This should be done in
accordance with all local code regulations. Make sure all connections are secure.
Warning! The exhaust port can become extremely hot. Do not touch. Avoid contact between the
port and flammable materials. Do not vent exhaust against combustible surfaces, areas, or
materials.
BLUE M ELECTRIC
A UNI T OF GE NERAL S IGN AL
®
STABIL-THERM
OVER
TEMP
OUTPUT
LATCH ED
UNDER
TEMP
1-800-448-9358
LIMIT
RESET
FM
on
off
EXHAUST MODIFICATIONS
Intake
Exhaust Hood
EXHAUST PORT with GUARD