Metalyser
®
Portable HM1000
Instruction Manual
v 5.0
22 www.trace2o.com
Obtaining
Results
Running
Analysis
Conditioning
5.4. STANDARD ADDITION
ANALYSES
OVERVIEW
Preforming
a
Standard
Addition method
analysis
consists of Plating
running Analysis
and
obtaining
a
result. The following
main steps will assist you in
understanding
how
to perform a
Standard
Addition based test.
Wipe & polish
WE
x
Pour appropriate Plating
Solution
into the SAB & attach sonde
Run Mx
Conditioning
on the Metalyser
®
Return
Mx
Conditioning Solution
to the bottle
Tip HMx
Buffer
into the
SAB
Add
sample
water to
SAB
Select appropriate
test
method
Run
Condition Electrode
on the Metalyser
®
Run Mx
Standard
Addition on the Metalyser
®
Add
Standard
to SAB when prompted
Continue Standard
Addition run on Metalyser
®
Save
/ discard / download results
Remove
SAB.
Rinse electrodes
& sonde
head
Where
‘WEx’, use
WE1
or
WE2
for your
required analysis
Where
‘Mx’, use M1, 2, 3, 4 or 5 for your
required analysis