9. SPECIFICATIONS
Metalyser
®
HM1000 Handheld Unit
Environment Water Ingress: IP67
Operating temp: -20ºC to +70ºC
Storage temp: -20 ºC to + 70 ºC
Connections 1 x waterproof Mini USB port, 1 x 12 Way Multipole
Connector.
Display 128 x 128 pixel monochrome backlit LCD.
Power supply 4.5 5V DC, 700mAh
Battery 3.3V Li-ion 2200mAh rechargeable.
Parameter
Lower
Limit
Upper Limit*
Arsenic (III)
As
(III)
5
500
Total
Arsenic
As (III+V)
10
500
Copper
Cu
5
500
Mercury
Hg
5
500
Cadmium
Cd
3
500
Lead
Pb
5
500
Zinc
Zn
5
500
*Upper
limit
fixed at
500ppb.
If
the user requires values higher than this then a
dilution
will
be
needed
Lower limits achievable
using the
Handset
at 60 seconds deposition.
Increased accuracy
can be
achieved
by setting the
deposition
time to 120 seconds
Variability
at 20ppb ± 5% with
Metaware
software, ± 7.5% with handheld
Accuracy
dependent on
element measured, sample
matrix and type