EasyManuals Logo

Trace2O METALYSER HM1000 User Manual

Trace2O METALYSER HM1000
29 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #28 background imageLoading...
Page #28 background image
28 www.trace2o.com
Metalyser
®
Portable HM1000
Instruction Manual
v 5.0
9. SPECIFICATIONS
Metalyser
®
HM1000 Handheld Unit
Environment Water Ingress: IP67
Operating temp: -20ºC to +70ºC
Storage temp: -20 ºC to + 70 ºC
Connections 1 x waterproof Mini USB port, 1 x 12 Way Multipole
Connector.
Display 128 x 128 pixel monochrome backlit LCD.
Power supply 4.5 5V DC, 700mAh
Battery 3.3V Li-ion 2200mAh rechargeable.
Parameter
Lower
Limit
Upper Limit*
Arsenic (III)
As
(III)
5
500
Total
Arsenic
As (III+V)
10
500
Copper
Cu
5
500
Mercury
Hg
5
500
Cadmium
Cd
3
500
Lead
Pb
5
500
Zinc
Zn
5
500
*Upper
limit
fixed at
500ppb.
If
the user requires values higher than this then a
dilution
will
be
needed
Lower limits achievable
using the
Handset
at 60 seconds deposition.
Increased accuracy
can be
achieved
by setting the
deposition
time to 120 seconds
Variability
at 20ppb ± 5% with
Metaware
software, ± 7.5% with handheld
Accuracy
dependent on
element measured, sample
matrix and type

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Trace2O METALYSER HM1000 and is the answer not in the manual?

Trace2O METALYSER HM1000 Specifications

General IconGeneral
BrandTrace2O
ModelMETALYSER HM1000
CategoryMeasuring Instruments
LanguageEnglish