14 Programming flow chart
5.1 Program detail record
If not powered for more than 10 seconds the programmed detail & Accumulated Total will
be lost from the processor memory, it is advisable to record your programmed detail
below. Specific instructions on changing the battery without loss of program detail are
given at clause 2.4, page 5.
6. TROUBLESHOOTING
# No display.
Check position of the battery jumper J2 ( see interface board layout page 10 ) & check
battery contact connections. Replace battery.
# Display ghosting under external power.
The LCD display will “ghost” when the instrument is powered from an external 24Vdc
power supply, this is corrected by including two ¼ watt resistors (1.2K & 3.3K) on the
input power source as shown below.
Troubleshooting 15
# Display shows software number at all times.
example
The instrument has not been fully programmed after power up, enter the program mode &
enter program parameters, be sure to exit the program mode before the 4 minute no data
entry time out or the instrument will revert back to model number display.
# Display ghosting under external power.
The LCD display will “ghost” when the instrument is powered from an external 24Vdc
power supply, this is corrected by including two ¼ watt resistors (1.2K & 3.3K) on the
input power source as shown below.
# Scaleable pulse output counts on after flow has stopped.
The scaled pulse output has exceeded its output limit of 8Hz. Allow the memory buffer to
catch up or increase the pulse value - number of litres etc. per pulse, (clause 4.5, p12.)
# Display shows random characters.
The instrument may have taken an electrical “hit”, press the processor reset button (p10).
Your back up 4 digit PIN number is 0220
0220 is referenced to the base model BT, “B” is the 2
nd
(02) & T” the 20
th
letter of the alphabet.
Cut from manual for increased security
Factory default settings
Pencil your program details here
K =
User selected PIN No.
K-factor (scale factor)
Engineering units
0 0.0 0.00 0.000
Decimal for Accum. total
Output pulse value
Decimal for reset total
0000
litres
1.000
0.0
0.0
0001.0
0 0.0 0.00 0.000
+Vdc
+∏
-gnd
4
5
6
BT11
ON
1 2 3
gnd
+
-
1
2
3
24Vdc
external
power
-
+
3.3K
1.2K