20
Set Infusion Parameters
Introduction to Infusion Parameters Setting
(1) The drug information can be displayed in the infusion running interface only when the drug
library is under active state.
(2) ClickSettings icon in the main interface to enter sub-menu, find Drug Library menu
item, click to enter then select drug. To remove drug information from running interface go back
to Drug library and select “none”. Please refer to Drug Library on Page 21 for more info.
(3) For both the rate set in infusion parameter and the rate calculated by the system, the range is
the system default flow rate of the current working syringe specification.
(4) If the user didn’t set a VTBI (Volume to be infused), this means the syringe will infuse until
empty. Infusion Parameters Setting Range
Infusion Parameter Parameter Range
VTBI 0-9999.99ml
Rate
(0.1-100) ml/h for 3ml syringes
(0.1-150) ml/h for 5/6ml syringes
(0.1-300) ml/h for /12 syringes
(0.1-600) ml/h for 20ml syringes
(0.1-900) ml/h for 30/35ml syringes
(0.1-1500) ml/h for 60ml syringes
Time 1min-99hrs59min
Weight (Body weight) 0.1-300kg
Active Agent (Drug mass) 0.01-99999.99
Conc. Unit (Concentration
unit)
ngmcgmggUkUIUIE
mmolmolkcal
Volume (Fluid amount) 0.01-9999.99ml
Dose rate 0.1-9999.99
Dose rate unit
Unit (/kg)/min, Unit(/kg)/h, the Unit is
Conc. unit