22
7. System Settings
Settings
Click『Settings』icon in the main interface to enter into parameters setting interface.
Drug Library
Click on the preset drug name, the selected drug will be reflected in infusion mode parameters.
To remove drug name from infusion mode parameters, select “none” in drug library. To Edit
drugs, go to Settings – Drug Library – DrugLib maintenance – Enter Password 8888 –
Edit/Add/Delete 2 different groups of drugs. Default Groups labeled “Commonly Used” and
“Others”. There are 32 total spots available for drugs, 16 in each group.
KVO Rate
Click『KVO rate』, input the numerical value, after confirming, click『OK』.
Bolus Rate
Set the default Manual Bolus rate under Settings.
Commonly used syringe brand
For the pre-installed syringe brands, after installing the syringe, click『Commonly used syringe
brand』to enter into the syringe brand selecting interface. Then click one of the pre-installed
syringe brands.
● The syringe of a different brand may possibly cause flow rate deviation. When using,
please confirm if the displayed information in the interface is accordant with the actual syringe
brand in use.
Occlusion Pressure
Click 『Occlusion pressure』to enter into occlusion level setting interface. Move the long box to
the preset level, after confirming, click 『OK』.
The higher the level, the higher the occlusion level, it is suggested to select a suitable occlusion
pressure according to actual requirement.