28
9. Maintenance
If pump displays the following: “Scheduled Maintenance Recommended Contact Triumph
Medical” this means your pump is do for its yearly calibration. Please proceed to
Triumphmed.com and fill out the Veterinary RMA form located at the top of the home
page.
Cleaning, disinfecting and sterilizing
Warning: Please cut off power supply and unplug the DC /AC power wire before
cleaning the equipment.
During cleaning and disinfecting, please keep the equipment horizontal and
upwards to protect the equipment and accessories from fluid.
Cleaning
(1) The daily maintenance consists of cleaning the shell and pump body. It is inevitable that
fluid/drug may flow in the equipment during infusion. Some fluid/drug may corrode the
pump and cause a fault. After infusion, please clean the equipment, wipe it with moist and
clean soft fabric. Let the pump dry.
(2) When cleaning the equipment interface, please wipe it with dry and soft fabric. Confirm the
interface is dry before using.
(3) Please do not soak the equipment in water. When fluid splashes on the equipment, please
check if it works normally even though the equipment has a certain waterproof function.
Perform insulation and electric leakage test if needed.
Disinfecting
(1) Disinfecting may possibly cause harm to the equipment. It is recommended to disinfect the
equipment only if it is needed.
Please disinfect the equipment with common disinfecting agent such as 50% sodium
hypochlorite, cidex 2% glutaraldehyde + activating agent, 70% ethanol, 70% isopropyl
alcohol and so on. Please follow the instructions of the disinfecting agent.
(2) After disinfecting, wet soft fabric with warm water, wring the excess liquid from the fabric,
and then wipe the equipment with it.