29
(3) Do not sterilize the equipment with a high-pressure steam sterilizer. Do not dry the
equipment with a dryer or similar product.
Warning: ● Please do not adopt Cidex OPA orthophthalaldehyde, methyl ethyl ketone or
similar solvent, otherwise, it may corrode the equipment.
Repair
Please contact Triumph Medical (888-388-3344 ext 3) to repair if there’s any fault. Do not
disassemble and repair the equipment.
Alarm and Solution
Alarm Type
Alarm
Level
Reason Solution
VTBI near
end
Low
During infusion, the remaining
time of preset value reaches or is
less than the set nearing
completion time
This alarm can’t be eliminated,
and waits till infusion completes
Syringe near
empty
Low
The syringe is near empty status
which is calculated by checking the
liquid medicine remaining in the
syringe by current flow rate.
This alarm cannot be
eliminated, must wait till
syringe is empty.
VTBI infused High
The preset value infusion
Completion
Press【Stop】
Syringe
empty
High
The liquid medicine in the syringe
is empty.
Press 【Stop】 button to stop the
alarm
Pressure near
threshold
Middle
Pipeline pressure increases close to
the preset blocking level.
Check the connection of the
pipeline, press 『OK』button to
continue infusion
Pressure drop Middle
When the pipeline pressure is
high,the pressure suddenly
decreases.
Check the connection of the
infusion pipeline, press 『OK』
button to continue infusion
Pressure high High 1. Line occlusion during infusion
Click 『 Mute 』 to silence,
manually remove the reason of
occlusion, Press 【Start】button
to continue infusion