Messages
Display Reason Action
Invalid
Hematocrit
Repeat with new test strip,
using capillary whole blood
from the nger or forearm.
If error persists, call
1-800-803-6025, available 24
hours a day, 7 days a week.
Temperature
Error
Too Cold/
Too Hot
Move meter and test strips
to area between 41°F-104°F;
wait 10 minutes for system
to reach room temperature
before testing.
Sample Not
Detected or
Using Wrong
Test Strip
Retest with new
test strip and larger sample.
Used Test Strip,
Test Strip outside
of vial too long,
Sample on top of
Test Strip.
Repeat with new test strip.
Make sure Sample Tip of test
strip touched top of sample
drop. If error persists, call
1-800-803-6025, available 24
hours a day, 7 days a week.
Meter
Error
Call 1-800-803-6025,
available 24 hours a day,
7 days a week.