7. ERROR DISPLAY
Error Message
Description
Solution
-----
Maximum load exceeded
Unload or reduce weight
Err 1
Incorrect date
Enter the date by using format
“yy;mm:dd”
Err 2
Incorrect time
Enter the time by using format
“hh:mm:ss”
Err 4
Zero setting error
Zero setting range exceeded due
to switching on.(4%max)
Make sure platform empty.
Err 5
Key board error
Check the keys and connecter.
Err 6
A/D value out of range
Make sure platform empty and
check the pan is installed proper.
Check the load cell connectors.
Err 9
Unstable Reading
Check any air variation, vibration,
RF noise and touching some
where.
Check the load cell and
connecters.
Err 17
Tare out of range
Remove the load and restart scale
again.
--ol--
Over range
Remove the load.
Re calibrate
Fai l h /
fai l l
Calibration Error
Re calibrate
Err p
Printer error
Check the printer and settings
Ba lo /
lo ba
Battery low
Re charge battery, check the
voltages.