Operation 3-41
6. Report will show the title you provided.
Field
Description
Zone
Select the zone from the dropdown list.
Standard
Select the standard from the dropdown list.
Class
Select the class from the dropdown list.
Use Start Date
Exclude samples collected before a specified date.
Use End Date
Exclude samples collected after a specified date.
Recipe
Check to include the recipe details on the report.
Exclude Samples
button.
Tap Exclude Samples button to display form that allows
you to enter IDs of samples to exclude and IDs of
replacement samples (if required by standard).
When you close the Exclude Samples form, the excluded
samples will be listed in the listbox on the Reports page.
Generate
Press to begin generating a report that you can view on-
screen or print.
The “Exclude Samples” form shown below is displayed when you click
the Exclude Samples button. Enter the ID of a sample to exclude in the
left column and enter the ID of a sample to replace it (if required by the
selected standard) in the right column. The software allows you to enter
up to five excluded and five replacement samples for ISO 14644-1:2015.
If the standard (e.g., ISO 14644-1:1999) allows only one excluded
sample and no replacements, only the first edit box in the left column will
be enabled and the replacement boxes will all be disabled.