55
3. Choose a location on the respirator that is in front of the person's
nose/mouth region. Avoid seams and folds in the respirator.
4. Pierce the respirator at the selected location, using the piercing
tool with the loaded sampling probe. BE SURE TO PIERCE THE
RESPIRATOR FROM THE INSIDE!
5. Push the sampling probe through the respirator until the end of
the tool point is visible from the outside of the respirator. Leave a
few millimeters of the tool point visible.
6. Place the push nut on the anvil tool with the "dished" side up. A
magnet prevents the push nut from falling off.
7. Position the anvil tool, with the loaded push nut, over the
protruding point. The respirator should now be pinched between
the two tools.
Sampling Probe
Push Nut
Positioned on
Anvil Tool
Sampling Probe
Pushed Through
Mask