Transient Response: < 20 µsec to within 50 mV of setting for 90% load change.
Temperature Coefficient: Typically < 100 ppm/
o
C.
Metering
(PL330QMT
only):
Meter Type:
3·75 digit (4095 count) with 12·5mm (0·5") LEDs. Reading rate 4Hz.
Meter Resolution:
Meter Accuracy:
10 mV and 10 mA
0·5% of reading + 1 digit.
GENERAL
AC Input Voltage: Internally set for 110, 120, 220, 230 or 240VAC 50/60 Hz; operating range
±10% of setting. Installation Category II.
Power Consumption: Single Dual Triple
15V / 4A or 30V / 2A: 160VA 320VA 400VA
30V / 3A: 250VA 500VA 600VA
Operating Range: 5
O
C to 40
O
C, 20% to 80% RH.
Storage Range: -20
O
C to +60
O
C.
Environmental: Indoor use at altitudes up to 2000m, Pollution Degree 2.
Weight: Single Dual Triple
15V / 4A or 30V / 2A: 5.0kg 9.5kg 12.5kg
30V / 3A: 6.0kg 12.0kg 15.0kg
Size: Single Dual Triple.
155 mm wide 350 mm wide 425 mm wide
All units 170mm high and 265mm deep, except PL330 versions 300mm
deep.
Safety: Complies with EN61010-1.
EMC: Complies with EN61326.