EasyManua.ls Logo

UNI-T UPO3000E Series - Delay Trigger Setup

UNI-T UPO3000E Series
94 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
UNI-T
UPO 3000E&UPO2000E Series Oscilloscope Programming Manual
Return format:
Query return 6.000e001unit is Hz.
For example:
:FUNC:SOUR1 CHAN1 set channel 1 as source 1
:FUNC:FILT:FREQ:LOW 60Hz set the low frequency limit of filter as 60Hz
:FUNC:FILT:FREQ:LOW? query return 6.000e001
:FUNCtion:LOGic:INVert
Command format:
:FUNCtion:LOGic:INVert {{1|ON}|{0|OFF}}
:FUNCtion:LOGic:INVert?
Functional description:
The switch setting of logic inversion function.
Return format:
Query return 1 or 0it presents ON or OFF.
For example:
:FUNCtion:LOGic:INVert OFF turn off inversion function
:FUNCtion:LOGic:INVert? query return 0turn off inversion function
:FUNCtion:EXPRession
Command format:
:FUNCtion:EXPRession <expression>
Functional description:
Use free combination expression to mathematical calculation.
Expression format see Advance item in MATH manu<expression> is belong to ASCII character
string parameter.
For example:
:FUNCtion:EXRP "CH1*CH2" channel 1 multiply channel 2