Check Is there anything obstructing the platter?
Action Remove all obstacles.
Syndrome Weighing result is not accurate
Check Is the scale placed in a level condition?
Action Adjust the adjustable feet to a level
condition.
Check Is the scale affected by airflow, vibration
or RFI?
Action: Place the scale away from all disturbances.
Check Is the scale calibrated correctly?
Action Contact your dealer.
Check Is the correct weight unit employed?
Action Refer to 5.3 for weight unit setting and
usage.
Syndrome Display shows 00000 after counting down
Check Is the scale overloaded?
Action Remove some loads from platter.
Check Is the stainless steel platter cover placed
on the platter?
Action Place the stainless steel platter cover
onto the platter.