4. INITIAL SETUP
4.1 PLACING THE SCALE
In order to obtain an accurate weighing result, the scale
must be placed on a strong and level surface.
If required, adjust the adjustable feet (FIG 2 F)
underneath scale to obtain a level position. Refer to
the bubble indicator for leveling condition.
4.2 INTERNAL SETTING
4.2.1 Display Segment Check
User can enter this function to check to see if the display
and backlight (if purchased) are functioning properly.
a. Scale is off
b. Press and hold TARE, then press ON/ZERO
c. Scale displays F.1
d. Press TARE once
e. Scale displays F.2
f. Press MODE and all segments will light up to allow
user to check the condition of display
g. Press TARE for next function setup or ON/ZERO to
restart the scale