Figure 11 RFL100 Properties in viewLinc
Local alarms on the RFL100 are also managed remotely, but in a dierent way. See Alarm
Indicators (page 11).
viewLinc automatically issues calibration reminder notifications at 3 months and 1
month before to the due date, and again on the due date.
4.2 Releasing RFL100 from a viewLinc Monitoring
System
Pen or a small screwdriver
After RFL100 Data Logger has been accepted to a viewLinc Monitoring System by a viewLinc
Enterprise Server administrator, it will not connect to any other system unless
first released
from its current system. Follow this procedure to release the data logger using its physical
Release button. Alternatively, you can release it remotely from viewLinc Enterprise Server
software.