8. Select menu item [3] Point 1 and press the Set button. The meter now shows the
currently measured value of the selected parameter. Set the reference value using the
arrow buttons and press the OK button.
The correction to the measurement at point 1 is now shown in the text for menu item [3]
Point 1. If you are only doing a 1-point calibration, skip to step 11.
9. Place the probe in the second reference environment (second calibration point). Wait 20 –
40 minutes for the reading to stabilize.